Catalog Number 6605500
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 09/29/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A field service engineer (fse) evaluated the analyzer on 09/29/2016, and indicated that the customer resolved the leak prior to the on-site service visit.The residual fluid on the sample tubes was attributed to the black plastic cap on port number 3 of the vic (vacuum isolator chamber) which fell off from the port, causing the vacuum to the probe wipe rinse block to drop, contributing to the leak.The fse inspected the analyzer and verified the plastic cap was properly reseated by the customer.(b)(4).
 
Event or Problem Description
A customer reported finding an uncontained leak of about 5cc of fluid on top of the sample tubes of a coulter ac t diff 2 analyzer while processing samples.The customer was wearing personal protective equipment (ppe) consisting of gloves and a laboratory coat when the leak was found, and there was no report of injury or biohazard exposure to open wounds or mucous membranes.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER AC·T DIFF 2 ANALYZER
Common Device Name
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
MDR Report Key6048968
Report Number1061932-2016-00879
Device Sequence Number359438
Product Code GKZ
UDI-Device Identifier15099590264796
UDI-Public(01)15099590264796(11)NO-DATA
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K990352
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/29/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number6605500
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2016
Initial Report FDA Received Date10/21/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-