• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MARVAO MEDICAL NEXSITE STEPPED TIP; HEMODIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MARVAO MEDICAL NEXSITE STEPPED TIP; HEMODIALYSIS CATHETER Back to Search Results
Catalog Number NEXHD1552801
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/03/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was indicated that the device will not be returned for evaluation because the device has been disposed; therefore an analysis of the complaint device could not be completed.A review of the manufacturing documentation confirmed that the lot met the specification requirements.The review found no deviations or non-conformances that would have contributed to the reported complaint.Taking into consideration the evaluation conducted and the details of the complaint, this investigation was assigned the most probable root cause of an anticipated procedural complication.A complaint with a most probable root cause classification of an anticipated procedural complication indicates that a device related root cause does not apply and the complaint is due to a known effect of the procedure.
 
Event or Problem Description
(b)(4): the nexsite device was successfully placed in this patient on (b)(6) 2016.Per the staff at the dialysis unit, the patient reported to dialysis on (b)(6) 2016 with a clotted access and mild poor flow.Activase was given to treat the event.The patient had his catheter removed due to this event on (b)(6) 2016 and the event had resolved on the same date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXSITE STEPPED TIP
Common Device Name
HEMODIALYSIS CATHETER
Manufacturer (Section D)
MARVAO MEDICAL
innovation in business centre
gmit, dublin road
galway, H91 D CH9
EI  H91 DCH9
MDR Report Key6050144
Report Number3008110587-2016-00034
Device Sequence Number13471711
Product Code MSD
UDI-Device Identifier05391525640020
UDI-Public(01)05391525640020(17)161101(10)42550
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K121933
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source study
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/10/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date11/01/2016
Device Catalogue NumberNEXHD1552801
Device Lot Number42550
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2016
Initial Report FDA Received Date10/24/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 YR
Patient Weight59
-
-