|
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
|
Manufacturing review: the device history records have been reviewed with special attention to the subassemblies, material review reports, raw material testing, manufacturing process, and quality control inspection.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Visual/microscopic inspection: the device was returned.The balloon size for this product was printed on the balloon hub of the catheter and identified the returned sample as a 6mm x 20mm balloon.No ruptures were noted to the balloon.Functional/performance evaluation: the patency was tested using an in-house.035¿ guidewire and passed without issue.The inflation hub was connected to an inflation device and an attempt was made to inflate the balloon with water.Upon inflation, water was observed to be leaking out the distal end of the strain relief.The strain relief was removed and a partial circumferential catheter shaft break was observed at the distal end of the y-hub.Sanding marks were noted on the catheter underneath the strain relief and at the location of the break.Medical records review: medical records were not provided; therefore, a review could not be performed.Image/photo review: images/photos were not provided; therefore, a review could not be performed.Conclusion: the investigation was confirmed for a partial circumferential catheter shaft break just distal to the y-hub, resulting in the reported leak.Per the drawing template, sanding marks shall not extend past the specified maximum sanding range indicated.Per the evaluation, sanding marks were identified past the hub (extending past the sanding range); therefore, excessive sanding of the catheter under the strain relief was the root cause for the partial break in the catheter.Labeling review: the current ifu (instructions for use) states: precautions: carefully inspect the catheter prior to use to verify that catheter has not been damaged during shipment and that its size, shape, and condition are suitable for the procedure for which it is to be used.Do not use if product damage is evident.Use of the ultraverse 035 pta dilatation catheter: position the balloon relative to the lesion to be dilated, ensure the guidewire is in place, and inflate the balloon to the appropriate pressure.
|