| Catalog Number |
6605632 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
10/03/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A field service engineer (fse) evaluated the instrument on 03-oct-2016.The fse found that the stabilyse tubing had disconnected from the diff mix chamber (vc25) causing the reagent to be dispensed onto the diff chamber module.The fse replaced the stabilyse input tubing to the side port of the vc25 and secured the tubing onto the port resolving the issue.The fse replaced a noisy diff mix motor assembly as a preventative measure.The repairs were verified per established service procedures.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported an unspecified volume of stabilyse leaked from the coulter lh 750 hematology analyzer.The customer was wearing personal protective equipment (ppe) consisting of a laboratory coat, gloves and glasses at the time of the event and there was no report of injury or biohazard exposure to open wounds or mucous membranes.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection to the event.There was no impact to patient results and controls.
|
| |
|
Search Alerts/Recalls
|