| Catalog Number |
6605632 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
10/02/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) evaluated the instrument.The fse found that tubing at the blood sampling valve (bsv) was crimped.The fse cleared reset the tubing, which repaired the leak.The repairs were verified by the fse.(b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported a leak from the coulter lh 750 hematology analyzer.The volume of the leak was about 5 ml and was not contained within the instrument.The customer was wearing personal protective equipment (ppe) consisting of gloves and a laboratory coat.There was no report of injury or biohazard exposure to open wounds or mucous membranes.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
|
| |
|