AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012452-08
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/06/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
An unused/sterile 3.75 x 8 mm nc trek balloon dilatation catheter was returned related to a case which had reported resistance removing the protective sheath.Returned device analysis confirmed resistance during removal of the protective sheath from this device.The protective sheath could not be removed.
 
Additional Manufacturer Narrative
(b)(4).The device returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda [medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.The abbott internal recall number is 2024168-3/14/2017-002-r.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6057759
Report Number2024168-2016-07312
Device Sequence Number863208
Product Code LOX
UDI-Device Identifier08717648152016
UDI-Public(01)08717648152016(17)190531(10)60603G1
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K110134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Remedial Action Recall,Other
Type of Report Initial,Followup
Report Date (Section B) 03/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Catalogue Number1012452-08
Device Lot Number60603G1
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/20/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 10/06/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/26/2016
Supplement Report FDA Received Date03/31/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-