| Catalog Number |
482391312S |
| Medical Device Problem Codes |
Fracture (1260); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/28/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the driver and tap damaged during case.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is pending due to the return of the device.Additional supplemental will be completed upon closure of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the driver and tap damaged during case.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection;device history review; complaint history review; risk assessment; results: the device was inspected and it was found that one of the tip prongs was fractured off.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Conclusion: the most likely root cause is wear due to the device's extended use.
|
| |
|
Event or Problem Description
|
|
It was reported that the driver and tap damaged during case.
|
| |
|