| Catalog Number |
723585 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
10/07/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A field service engineer (fse) evaluated the instrument on 10/07/2016 and found a leak from check valve (cv 8) letting clenz bubbles out of the white blood cell (wbc) bath.The fse replaced the check bath resolving the leak.The fse also replaced pinch valve (vl10) which was loose and broken but did not malfunction and the aperture o-rings as a preventative measure.The instrument ran without leaks and all repairs were verified per established service procedures.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported a leak of blood-diluent and cbc lyse fluid of approximately < 1.0 ml in volume which was contained within the coulter lh 780 hematology analyzer.The customer was wearing personal protective equipment (ppe) consisting of a laboratory coat and gloves at the time of the event.There was no report of injury or biohazard exposure to open wounds or mucous membranes.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.
|
| |
|