Model Number MICL12.6
Medical Device Problem Codes Positioning Failure (1158); Torn Material (3024)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/01/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual inspection of the returned product found one haptic torn, with pieces torn off and missing.The lens was returned in liquid.(b)(4).
 
Event or Problem Description
The reporter stated the surgeon was inserting a 12.6mm micl12.6 implantable collamer lens, -8.5 diopter, and noted the lens was damaged.The lens was partially inserted and was removed with no patient injury.The backup lens was implanted.The reporter stated the lens may not have been pulled down the cartridge barrel enough and the foam tip plunger caught and damaged the lens.The reporter stated the cause of the event was due to user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key6064990
Report Number2023826-2016-01560
Device Sequence Number16360811
Product Code MTA
UDI-Device Identifier00841542102995
UDI-Public(01)00841542102995(17)180430
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/13/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date04/30/2018
Device Model NumberMICL12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2016
Initial Report FDA Received Date10/28/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL SFC-45 FP - LOT NUMBER UNK; FOAM TIP PLUNGER MODEL FTP - LOT NUMBER UNK; INJECTOR MODEL AND LOT NUMBER UNK
Patient Age36 YR
-
-