| Catalog Number |
03822601 |
| Medical Device Problem Code |
Disassembly (1168)
|
| Health Effect - Clinical Codes |
Injury (2348); No Code Available (3191)
|
| Date of Event |
10/04/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that; rods slipped through the tulip heads of the pedicel screws.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, device history review, complaint history review, risk assessment.Result: inspection of the returned device did not reveal any visual non-conformances.No relevant manufacturing issues were identified as all units met stryker specifications.Conclusion: the most likely cause of the customer reported event is the high activity level of the patient with the over-tightening likely contributing to the event.
|
| |
|
Event or Problem Description
|
|
It was reported that; rods slipped through the tulip heads of the pedicel screws.
|
| |
|