| Catalog Number |
700-7177-001 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/04/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) was at the customer site and observed that there was a leaking tubing at the end of the color gravity module (cgm) assembly.The fse replaced the tubing at the cgm and confirmed there was no more leaking.The repairs were verified per established service procedures.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported a contained leak on their ichem velocity automated urine chemistry system.The customer also stated they were experiencing chem na and short sample errors on all patient samples and controls.The customer was wearing personal protective equipment (ppe) at the time the leak was observed.There were no reports of exposure to open wound or mucous membranes.No one sought medical attention.Erroneous patient results were not reported out and there was no change or effect to patient treatment in connection to the event.
|
| |
|