MAUDE Adverse Event Report: AV-TEMECULA-CT RX VIATRAC 14 PLUS PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT RX VIATRAC 14 PLUS PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number 1008193-20
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/05/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no similar incidents from this lot.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The investigation was unable to determine a conclusive cause for the reported difficulties.The 4.0 x 20 mm x 135 cm viatrac plus referenced is being filed under a separate medwatch report.
 
Event or Problem Description
It was reported the procedure was to treat a lesion in the common carotid artery.During preparation of the 5.0 x 20 mm x 135 cm viatrac plus balloon catheter, when pulling negative during preparation, there was air coming into the syringe due to a hole that was noted on the shaft; therefore, the device was not used.The 4.0 x 20 mm x 135 cm viatrac plus balloon catheter was not able to hold negative pressure, there was air coming into the syringe; therefore, the device was not used.The 4.0 x 30 mm x 135 cm viatrac plus had some difficulty pulling negative, but was able to be prepped.The device was advanced to the target lesion, the balloon inflated and deflated successfully.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX VIATRAC 14 PLUS PERIPHERAL DILATATION CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6074249
Report Number2024168-2016-07511
Device Sequence Number1602446
Product Code LIT
UDI-Device Identifier08717648056987
UDI-Public(01)08717648056987(17)180731(10)6082361
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/02/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2018
Device Catalogue Number1008193-20
Device Lot Number6082361
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2016
Initial Report FDA Received Date11/02/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-