• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY ICL (IMPLANTABLE COLLAMER LENS); INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/10/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Lens not returned.
 
Event or Problem Description
The reporter stated that a 13.2mm micl13.2 lens, -12.00 diopter, tore during loading on (b)(6) 2016.There was no patient contact and a backup lens was used.The cause of the tear is unknown.
 
Additional Manufacturer Narrative
Device evaluation: product evaluation found that the lens was returned in liquid in the lens vial.Visual inspection found the haptic torn with a piece missing.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICL (IMPLANTABLE COLLAMER LENS)
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key6074378
Report Number2023826-2016-01570
Device Sequence Number549681
Product Code MTA
UDI-Device Identifier00841542103336
UDI-Public(01)00841542103336(17)180131
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/24/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date01/31/2018
Device Model NumberMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 10/24/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/02/2016
Supplement Report FDA Received Date11/21/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL UNK, LOT NUMBER UNK; INJECTOR MODEL UNK, LOT NUMBER UNK
Patient Age24 YR
-
-