| Model Number |
ZLB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Halo (2227); No Code Available (3191)
|
| Date of Event |
10/06/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zlb00 18.0 diopter intraocular lens (iol) was explanted from a male patient's right eye due to halos.The zlb00 was explanted and replaced with a zct225 18.0 diopter iol.Lens exchange was successful with no incision enlargement, no vitrectomy and no sutures required.No post op patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device was returned to the manufacturer.Device inspection was performed and the lens was cut in half, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The device issue could not be confirmed in the returned sample.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no similar complaints were received for this order number to date.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the intraocular lenses.Based on the investigation results there is no indication of product quality deficiency.All pertinent information available to the manufacturer has been submitted.
|
| |
|