| Model Number |
E2750 |
| Medical Device Problem Code |
Device Remains Activated (1525)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/06/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Covidien reference#: (b)(4).Date of initial report: 11/03/2016.Evaluation of the incident pencil confirmed the customer's report.The cause is due to an assembly related failure.
|
| |
|
Event or Problem Description
|
|
The customer reported that during a total laparoscopic colectomy procedure, the coag button did not return to the off position.There was no patient harm.A new device was opened to continue the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date of initial report: (b)(6) 2016.Date of follow-up report: 02/13/2017.The customer reported the device had a latch-on condition.The reported condition was not confirmed.An additional fault was found during the investigation.The additional finding did not contribute to the customer¿s complaint.The investigation found the rocker switch activation force to be out of specification.The width of the rocker was found to be larger than the width of the rocker window in which it moves causing friction and an out of specification activation force.The width of the rocker window is formed by the ultrasonic welding of the two body housing components.The root cause of the activation force being out of specification was caused by interference between the rocker component and the rocker window in which it moves.Corrective action is being issued.
|
| |
|