| Catalog Number |
1012448-20 |
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
10/17/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.Investigation is not complete.A follow up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that during device preparation, the 2.75x20 mm nc trek rx balloon dilatation catheter was removed without resistance from the hoop, and the mid-shaft separated.The device was not used and there was no patient involvement.There was no clinically significant delay in the intended procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Evaluation summary: visual inspection was performed on the returned device.The reported separation was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The investigation was unable to determine a conclusive cause for the reported shaft separation.
|
| |
|