• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; PLATE Back to Search Results
Model Number 25-398-31-09
Health Effect - Clinical Code Impaired Healing (2378)
Type of Reportable Event Serious Injury
Event or Problem Description
Two plates broke in half after implantation into the patient.Patients fracture did not heal.Plates were removed on (b)(6) 2016 and replaced.
 
Additional Manufacturer Narrative
Methods: visual inspection, stereo microscopic inspection and process evaluation based on the lot number.Results: tensile cracks were observed under the stereo microscope.Further observation determined there were no indications of material or manufacturing defects observed.The product history device records were also reviewed based on the lot number provided, and no abnormalities were found.Assessment conclusion: the results of the investigation conclude that the root cause for breakages was due to mechanical overload on the device.If further information can be gathered that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
PLATE
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM  78570
MDR Report Key6082918
Report Number9610905-2016-00041
Device Sequence Number13520761
Product Code JEY
UDI-Device Identifier00888118039750
UDI-Public(01)00888118039750
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K943347
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Model Number25-398-31-09
Device Lot Number32794064
Was Device Available for Evaluation? Yes
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/07/2016
Supplement Report FDA Received Date01/06/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age52 YR
-
-