MEDLINE RENEWAL MEDLINE RENEWAL REPROCESSED ACE+ SHEARS WITHOUT ADAPTIVE TISSUE TECHNOLOGY; ULTRASONIC SCALPEL, HARMONIC SCALPEL Back to Search Results
Model Number HAR36M
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/03/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device in question was not returned to medline renewal, however an unused, reprocessed har36m from the same lot that was still sealed in the original medline renewal package was returned for evaluation.Our investigation included both an inspection of the returned device, and a review of the device history record (dhr).During the investigation, the device was subjected to all inspection and testing procedures, and as a result, met all acceptance criteria.In addition, our review of the dhr reconfirmed that all processes were conducted as required, and that the device met inspection requirements prior to packaging and release.The reported issue could not be confirmed at this time.The customer confirmed that no adverse patient consequence or additional medical intervention was required as a result of the incident.However, due to the reported delay in the procedure to obtain a new device, medline renewal is filing this medwatch report.H3 other text : another device from same lot evaluated.
 
Event or Problem Description
Medline renewal received a report indicating that a reprocessed ethicon harmonic ace+ laparoscopic shears without adaptive tissue technology, model har36m, did not connect properly to the generator, and that the device would not activate during use.Another device was readily available to complete the procedure; therefore no patient harm or additional medical intervention was reported as a result of the incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE RENEWAL REPROCESSED ACE+ SHEARS WITHOUT ADAPTIVE TISSUE TECHNOLOGY
Common Device Name
ULTRASONIC SCALPEL, HARMONIC SCALPEL
Manufacturer (Section D)
MEDLINE RENEWAL
2747 sw 6th st.
redmond OR 97756
MDR Report Key6085810
Report Number3032391-2016-00040
Device Sequence Number774940
Product Code NLQ
UDI-Device Identifier10888277408708
UDI-Public10888277408708
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K151405
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 11/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date06/30/2019
Device Model NumberHAR36M
Device Lot Number307295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2016
Initial Report FDA Received Date11/08/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-