MAUDE Adverse Event Report: SYNTHES OBERDORF SAGITT SAW-ATTACHM LONG W/T-HANDLE F/TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES OBERDORF SAGITT SAW-ATTACHM LONG W/T-HANDLE F/TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.224
Device Problems Partial Blockage (1065); Material Discolored (1170); Overheating of Device (1437); Sticking (1597); Device Inoperable (1663); Incorrect Or Inadequate Test Results (2456); Appropriate Term/Code Not Available (3191)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was duplicated and confirmed.The assignable root cause was determined to be due improper handling, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from (b)(6) that the sagittal saw attachment device was stuck; and that it was not working.During service and evaluation, it was observed that the sagittal saw attachment device bearing did not function and the device was blocked.It was further determined that the device failed the following pre-tests: check of free movement, check the oscillation frequency and check the saw performance.During troubleshooting, it was identified that the needle bearing was fixed but black discoloration was found on the shaft due to overheating.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAGITT SAW-ATTACHM LONG W/T-HANDLE F/TRS
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ   CH4436
Manufacturer Contact
mark vornheder
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6088957
MDR Text Key59722338
Report Number8030965-2016-15503
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 08/22/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/09/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.224
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2016
Date Manufacturer Received10/28/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/26/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-