Model Number PE075F5
Device Problem Pacing Problem (1439)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/17/2016
Event Type  malfunction  
Manufacturer Narrative
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product a supplemental report will be sent with the investigation results.A device history record review was completed and documented that device met all specifications upon distribution.
 
Event Description
It was reported that the catheter was unable to sense an intrinsic electrical signal from the beginning of use.The catheter was replaced and the problem was solved.It is unknown if a problem with pacing was observed.Further detail could not be obtained.There were no patient complications reported.Patient demographic information requested but unavailable.
 
Manufacturer Narrative
One bipolar pacing catheter with attached monoject 1.3 cc limited volume syringe was returned for evaluation.Continuity testing confirmed a full open condition of the distal circuit.The proximal circuit was found to be continuous.A cut down of the catheter body was performed just proximal of the proximal electrode to expose the pacing leadwires.The distal leadwire was found to be broken around the solder joint.The balloon inflated clear and concentric with 1.3 cc air and remained inflated for 5 minutes without leakage.No visible damage or abnormality was observed from the balloon, windings or returned syringe.Balloon inflation test was performed using returned syringe with 1.3 cc air by holding the balloon under water.Visual examinations were performed under microscope at 20x magnification and with the unaided eyes.Customer report of pacing issue was confirmed.An investigation has been initiated to consider any potential manufacturing issues that may have contributed to this event and implement any necessary corrective actions.Swan-ganz pacing thermodilution (td) catheters serve as diagnostic and therapeutic tools in the management of critically ill patients.There are multiple failure modes that may require the exchange of a pacing catheter.Since proper functioning of the pacing catheter depends on the electrical continuity of its electrodes and internal wires, care should be exercised when handling the catheter.Stretching, kinking, or forceful wiping of the catheter may result in damage.After stable pacing has been confirmed, the proximal end of the catheter should be secured to the insertion site to prevent undue movement that could result in tip dislodgment and loss of capture, or catheter migration.Care should be taken not to kink the catheter body when securing it.In this complaint, it could not be determined if procedural factors or device handling may have contributed to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ BIPOLAR PACING CATHETER FOR FEMORAL INSERTION
Type of Device
PACING CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
one edwards way
irvine, CA 92614
9497564386
MDR Report Key6094352
MDR Text Key59720186
Report Number2015691-2016-03338
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/18/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/10/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/18/2018
Device Model NumberPE075F5
Device Lot Number60366158
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received11/10/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-