• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ATRIUM MEDICAL CORP C-QUR TACSHIELD; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRIUM MEDICAL CORP C-QUR TACSHIELD; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number 31633
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/08/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow up report will be submitted after the completion of the investigation into this event.
 
Event or Problem Description
During an open ventral hernia case there were two pieces of mesh in the sterile packet of a tacshield rather than one.
 
Additional Manufacturer Narrative
Engineering analysis: engineering reviewed the sterilization and device history records for the product in question and review showed that criteria was met.Engineering summary: the picture provided by the hospital shows a "spare" piece of tacshield after implanting one of the devices.The spare part was discarded rather than sent back for evaluation, so engineering cannot confirm the tacshield in the picture was the one scrapped in process.A review of past product complaints by the complaints department show that we have not had a complaint for multiple parts in the same packaging.Given the controls in place, this is likely an isolated incident that was not identified during the 100% inspection visual inspection just prior to placing product in a handling sleeve in process.As far as the assessment of whether or not the parts were sterile, engineering is confident that the two pieces of c-qur tacshield cited in the complaint were not a worst case challenge when compared to the validations that had been performed.Based on the assessment detailed in the investigation and the results from those sterilization validations, the parts are sterile and should perform as intended.Clinical evaluation: if there is a compromise in the integrity of the packaging of a sterile device, it would be noticed prior to being introduced to the surgical site.This may create a delay in the procedure while a second device is located and prepped.The complaint states that two pieces of mesh were found inside the sterile packet instead of just one.In the event that an additional piece of the product was added to the package after receipt at the facility the sealed pouch and the seal on the outer box would be broken.The instructions for use warn, 'this device is supplied sterile.Please inspect packaging to ensure it is intact and not damaged prior to use.'.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C-QUR TACSHIELD
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
ATRIUM MEDICAL CORP
40 continental blvd.
merrimack NH 03054
MDR Report Key6095395
Report Number3011175548-2016-00009
Device Sequence Number13793102
Product Code FTL
UDI-Device Identifier00650862316339
UDI-Public00650862316339
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K100076
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date08/11/2019
Device Model Number31633
Device Catalogue Number31633
Device Lot Number408265
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 11/08/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/11/2016
Supplement Report FDA Received Date12/28/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-