| Model Number |
73779 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/16/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Visual inspection of the returned screw extender revealed that both of the locking tabs have been completely removed/broken from the instrument.Both detached tabs were returned with the remainder of the instrument conforming the patient did not retain a foreign body.An evaluation of the of the device history records found no manufacturing, processing or design related irregularities.The instrument was properly manufactured and released in accordance with the device master record.Although the device was properly manufactured additional investigation was conducted to determine root cause.The analysis showed that the failure mode observed in this event was due to imparting excessive torsional loads to the mating instrument (reducer) during the rod reduction maneuver.Excessive load can be applied due to a tolerance gap condition between the reducer and extenders.The over-torque observation was corroborated by the two design interface analyses (dias) that were performed on the reducer/screw extender interfaces.
|
| |
|
Event or Problem Description
|
|
An international customer ((b)(6)) reported that the illico instrument broke during surgery.
|
| |
|