• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. FOUNDATION UNIPOLAR WITH NECK LENGTH SLEEVE; SLEEVE, UNIPOLAR, OFFSET NEUTRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. FOUNDATION UNIPOLAR WITH NECK LENGTH SLEEVE; SLEEVE, UNIPOLAR, OFFSET NEUTRAL Back to Search Results
Catalog Number 411-00-000
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 10/12/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having an acute infection; the surgeon removed and replaced the liner head and sleeve.
 
Additional Manufacturer Narrative
The reason for this revision surgery was an acute infection.The previous surgery and the revision detailed in this investigation occurred 20 days apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.There were no findings during this investigation that indicate that the reported device was the root cause or had a direct connection with the patient's infection.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post surgical instructions.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOUNDATION UNIPOLAR WITH NECK LENGTH SLEEVE
Common Device Name
SLEEVE, UNIPOLAR, OFFSET NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin TX 78758
MDR Report Key6096746
Report Number1644408-2016-00837
Device Sequence Number840366
Product Code KWL
UDI-Device Identifier00888912076692
UDI-Public(01)00888912076692
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K973614
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/23/2022
Device Catalogue Number411-00-000
Device Lot Number902B1938
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 10/12/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/11/2016
Supplement Report FDA Received Date01/04/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
497-36-000, LOT 638C1178; 931-36-752, LOT 761N1945
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
-
-