The patient underwent a generator replacement due to neos=yes on (b)(6) 2016.The patient's lead was replaced the same day for compatibility reasons.Prior to replacement surgery, system diagnostics confirmed system function.After surgery, two lead portions were received by the manufacturer for product analysis, however the electrode array was not included.The patient's lead was found to be fractured within the tubing and in the same area an abrasion in the outer and inner tubing was found.The abrasion was believed to be due to wear.Fluid was found within the lead inner tubing and this was attributed to the brasion.Regarding the lead fracture, two strands of the lead were fractured due to wear, the final strand of the lead was fractured due to rotational stress that likely occurred during product analysis continuity checks.Trauma or manipulation to the site of the abrasion and fracture was denied.No additional relevant information has been received to date.
|