| Catalog Number |
1012452-15 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/15/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).The device was received.The investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that during device preparation, resistance was met during removal of the protective balloon sheath and the device was not submerged to activate the coating.The device was taken to the lesion in the left anterior descending (lad) artery and was positioned, but failed to inflate.The device was removed and a different device was used in the procedure without issue.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Evaluation summary: the device returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda on march 17, 2017 [medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.The abbott internal recall number is 2024168-3/14/2017-002-r.
|
| |
|
Search Alerts/Recalls
|