Model Number PED-475-35 |
Device Problem
Unintended Movement (3026)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/13/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
The pipeline flex pushwire and microcatheter were returned for evaluation; however, the pipeline braid was not returned as it was implanted in the patient.The tip coil appeared to be stretched.In addition, the distal hypotube appeared to be stretched with the ptfe shrink tubing still intact.The pushwire was also found to be bent.The microcatheter appeared to be cut and the proximal segment of the catheter along with the hub was not returned.The microcatheter segment returned was noted to be flattened with accordion damage.No other anomalies were observed.Based on the analysis findings we are unable to definitively determine the cause for the reported event.It is possible that the reported moderate vessel tortuosity may have contributed to the resistance during retrieval issue; subsequently causing the pipeline flex delivery system and catheter to become damaged.However, the cause for resistance could not be determined.Additionally, the damages seen on the catheter body (flattening/accordioning), hypotube (stretching), distal and proximal wire (bending); it appears there was excessive force used (pushing and pulling).Per our instructions for use (ifu), the user should discontinue delivery of the device if high force or excessive friction is encountered during delivery.Identify the cause of the resistance and remove device and microcatheter simultaneously.Advancement of the ped against resistance may result in device damage or patient injury.Never advance or withdraw an intralumenal device against resistance until the cause of resistance is determined by fluoroscopy.If the cause cannot be determined, withdraw the catheter.Furthermore, do not use the pipeline flex embolization device in vessel diameters that are larger than the labeled diameter.Select an appropriately sized pipeline flex embolization device such that its fully expanded diameter is equivalent to that of the largest target vessel.An incorrectly sized pipeline flex embolization device may result in inadequate device placement, incomplete opening, or migration.All products are 100% inspected for damage and irregularities during manufacture.Mdrs related to this report: 2029214-2016-01011, 2029214-2016-01012.
|
|
Event Description
|
Medtronic received information that during treatment of an aneurysm located in the left posterior communicating artery, the first device (ped-425-35) landed at the desired location.However during planned percutaneous transluminal angioplasty (pta) the device foreshortened causing the neck to not be fully covered.In order to cover the neck completely, this (ped-475-35) a second device was placed.It was reported that the physician planned to place two pipeline devices.However, upon placing the second pipeline, the same microcatheter was used and the physician was unable to resheath, and the initial pipeline and the second pipeline were apart.The physician made sure the pusher and catheter were able to pass through the pipeline implanted, and tried to advance the guidewire.However, the physician had difficulty removing the pushwire of the pipeline.The physician then decided to cut the microcatheter, a few tenths of a cm from the proximal end, exposed the pusher and attempted removing from the patient but was unable to do so.The physician eventually removed the pusher in tandem with the microcatheter.The second pipeline was not implanted in the intended location and therefore a third pipeline was then implanted between the first two pipelines.The procedure was completed.No patient injury was reported.The aneurysm max diameter was estimated to be 10 mm.The aneurysm neck is unknown.The proximal landing zone was 4.8 mm and the distal landing zone was 4.1 mm.The aneurysm was located in the left internal carotid artery approximately 50 mm proximal of the pcom branch.The patient tortuosity was medium level.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|