• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT375
Medical Device Problem Codes Incorrect Measurement (1383); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/24/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was initially reported that the doctor likes the lens and has had success with other patients with the lens but was planning to explant the lens as the axis has been flipped on this male patients left eye, resulting in 1.5d of residual cylinder.He mentions his wound does not create a large astigmatism shift, but the pre-op pentacam may have shown some artifact astigmatism.He believes using the symfony toric calculator when picking the cylinder correction power would have been beneficial.He will replace this symfony toric with a tecnis toric lens.Further information confirmed the explant did occur due to the toric axis flipped.The physician picked the wrong cylinder correction power because he did not use the toric calculator.Additionally, the physician noted that the patient had a 20/20 refraction.The patient has recovered.No other information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation ¿ yes, returned to manufacturer on 12/16/2016.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturer.Visual inspection using 12x magnification and does not show anomalies.The condition of the lens is consistent with an explanted product.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no similar complaints were received for this order number to date.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the intraocular lenses.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6105094
Report Number9614546-2016-00606
Device Sequence Number10930777
Product Code POE
UDI-Device Identifier05050474606944
UDI-Public(01)05050474606944(17)210718
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date07/18/2021
Device Model NumberZXT375
Device Catalogue NumberZXT375U185
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 10/19/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/16/2016
Supplement Report FDA Received Date02/01/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-