| Model Number |
ZXT375 |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/24/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was initially reported that the doctor likes the lens and has had success with other patients with the lens but was planning to explant the lens as the axis has been flipped on this male patients left eye, resulting in 1.5d of residual cylinder.He mentions his wound does not create a large astigmatism shift, but the pre-op pentacam may have shown some artifact astigmatism.He believes using the symfony toric calculator when picking the cylinder correction power would have been beneficial.He will replace this symfony toric with a tecnis toric lens.Further information confirmed the explant did occur due to the toric axis flipped.The physician picked the wrong cylinder correction power because he did not use the toric calculator.Additionally, the physician noted that the patient had a 20/20 refraction.The patient has recovered.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation ¿ yes, returned to manufacturer on 12/16/2016.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturer.Visual inspection using 12x magnification and does not show anomalies.The condition of the lens is consistent with an explanted product.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no similar complaints were received for this order number to date.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the intraocular lenses.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
|
| |
|
Search Alerts/Recalls
|