ARTHREX, INC. BIO-COM INTERFERENCE SCREW, TAPERED TIP; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
AR-1360C |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/26/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No further patient information was provided at the time of this report or made available in response to follow-up communication.No additional adverse consequences have been reported from this event.This device is used for treatment.The device was requested for evaluation but was not returned, therefore the complainant's event could not be verified.The cause of the event could not be determined from the information available and without device evaluation.Device history record review revealed nothing relevant to this event.Complainant's event is typically caused by not inserting the implant co-axial to the bone tunnel, prying or leveraging while still loaded, improper bone preparation or flexing the joint during insertion.This is the (b)(4) complaint of this type for this part/lot combination.The potential cause(s) of this event will be communicated to the event reporter.If additional relevant information is received, a follow-up report will be submitted.Discarded by the facility.
|
| |
|
Event or Problem Description
|
|
It was reported that the 6mm x 23mm biocomposite interference screw was being used in a mpfl reconstruction procedure.The surgeon drilled the pilot hole using a reamer.When he went to insert the interference screw, the screw broke in half.Half of the implant remains in the bone of the patient, not proud.The surgeon decided to use a cortical button on the other side of the broken implant for a secondary fixation to complete the procedure successfully.No patient injury reported.
|
| |
|
Search Alerts/Recalls
|
|
|