Model Number CI-1400-01 |
Device Problems
Device Operates Differently Than Expected (2913); Electrical Shorting (2926)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/26/2016 |
Event Type
malfunction
|
Event Description
|
The recipient is reportedly experiencing open electrodes and a decrease in performance.Revision surgery will be scheduled.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
The external visual inspection revealed a kinked electrode wire at the small tubing of the electrode.The device passed the photographic imaging inspection.System lock was verified.The device passed all of the electrical and mechanical tests performed.This is an interim report.
|
|
Manufacturer Narrative
|
The external visual inspection revealed a kinked electrode wire at the small tubing of the electrode prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed all of the electrical tests performed.The device failed the residual gas analysis test.The scanning electron microscopy analysis revealed cracked conductive epoxy at the feedthru pin connections on some pins.The impedance measurement across these connections did not reveal any increased impedance.Dissection and scanning electron microscopy analysis of the electrode lead and array revealed broken electrode wires.This is not believed to be related to the return reason.The reported complaint of electrode opens could not be verified during this analysis.This device passed the electrical tests performed.However, this device had moisture that exceeded the residual gas analysis test limit.The source of the problem was a feedthru hermeticity issue from one feedthru vendor.A capa was implemented.Feedthru assemblies from this vendor are no longer used.This is the final report.
|
|
Search Alerts/Recalls
|