MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH FIBERSCOPE "CYF-5", EUROPEAN VERSION; CYSTOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH FIBERSCOPE "CYF-5", EUROPEAN VERSION; CYSTOSCOPE Back to Search Results
Model Number CYF-5
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/17/2016
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to olympus medical systems corp.But was returned to olympus (b)(4).Olympus sent the subject device to a third party laboratory for microbiological testing after reprocessing and the testing indicated no microorganisms growth for the subject device.The exact cause could not be determined at present.If significant additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that the subject device repeatedly tested positive for aerobic bacteria (1cfu-3cfu) during routine surveillance culturing by the facility.There was no report of infection associated with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERSCOPE "CYF-5", EUROPEAN VERSION
Type of Device
CYSTOSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer Contact
hiroki moriyama
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
8142642517
MDR Report Key6108382
MDR Text Key60084810
Report Number8010047-2016-10059
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Other Device ID Number04042761051453
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/20/2016
Initial Date FDA Received11/17/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-