Model Number N/A
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/14/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.
 
Event or Problem Description
Per the mw(b)(4), reported through the fda voluntary event reporting process, the manufacturer was informed that the guide wire was not removed after chest tube placement.The reporter alleged that the "tag that says "remove dilator after chest tube insertion" is not very noticeable or understandable." it was further reported to the manufacturer in response to inquiry's that: the drainage being performed was pneumothorax, no unintended sections of the device remained in the patient, an additional procedure was performed, the complained device was in place for 1 day, the chest tube was replaced with a different device and remained in for 2 days, the patient is doing well and no adverse events occurred due to this event.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the device history record, manufacturing instructions, quality control, and specifications of the returned device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THAL-QUICK CHEST TUBE SET
Common Device Name
GBX CATHETER, IRRIGATION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key6110276
Report Number1820334-2016-01337
Device Sequence Number1123628
Product Code GBX
UDI-Device Identifier00827002067868
UDI-Public(01)00827002067868(17)190330(10)6854808
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberC-TQTS-800
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 10/25/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/17/2016
Supplement Report FDA Received Date05/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age1 DA
-
-