| Model Number |
ZKB00 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
09/30/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a zkb00 20.0 diopter intraocular lens was explanted from the right eye of a female patient due to an unexpected post up refraction.Lens was replaced with the same model lens but with a 19.0 diopter.There was no incision enlargement, no vitrectomy, and no sutures required.No patient injury post-op and patient is doing well.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device was not returned to the manufacturer for evaluation.Device inspection could not be performed, therefore the reported complaint could not be verified.Manufacturing records were reviewed and the lens was manufactured according to specification.A search on complaints revealed no similar complaints were received for this production order number.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the intraocular lenses.Based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
|
| |
|