Catalog Number 490368
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/08/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The actual device involved in the reported incident was not returned for evaluation.Without the actual sample, a thorough sample analysis could not be performed and no specific conclusions could be drawn.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or nonconformances of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number or lot number.If additional pertinent information becomes available, a follow-up report will be filed.Note: this case is being filed retrospectively as a result of a review of recent customer complaint information.Based on additional information and details provided in another complaint case, it was determined that this case is reportable in accordance with the requirements of 21 cfr 803.B.Braun has conducted a retrospective review for all complaints of a similar nature in accordance with internal procedure (b)(4).
 
Event or Problem Description
As reported by the user facility: reports an iv was started with 500 ml 0.9% sodium chloride attached to a patient.The roller clamp was adjusted to a slow drip for kvo rate.Thirty minutes later, the bag was empty.The nurse tried adjusting the roller clamp on another set and noted it had to be rolled all the way to almost off to allow for a very slow drip rate.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV ADMINISTRATION SET
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
MDR Report Key6112798
Report Number2523676-2016-00729
Device Sequence Number13524937
Product Code FPA
UDI-Device Identifier04046964669927
UDI-Public(01)04046964669927(17)170131(10)0061419942
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K111016
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 04/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2017
Device Catalogue Number490368
Device Lot Number0061419942
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2015
Initial Report FDA Received Date11/18/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-