| Model Number |
800724-1 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Operational Issue (2914); Maintenance Does Not Comply To Manufacturers Recommendations (2974)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
10/27/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
During the repair on (b)(6) 2016, our field service technician noted that it appeared maintenance on the machine was performed improperly.The motor was returned and our engineer performed an investigation on the motor.
|
| |
|
Event or Problem Description
|
|
In support of the 2016 correction a field service technician was sent in to inspect the safety mechanism of two machines.He was unable to inspect one machine because the shaft would not spin and the machine would not travel up/down.Since this problem was not reported when the inspection was setup the technician did not have parts to repair the machine and would have to return to repair the machine.The repair was schedule and performed on (b)(6) 2016.During the repair it was determined that the screw motor had failed and the unit was supported by the safety nut.
|
| |
|