• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES C4119, 32CM GRASPER REPOS CART, 10/BX; NWV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES C4119, 32CM GRASPER REPOS CART, 10/BX; NWV Back to Search Results
Model Number C4119
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/25/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident device is anticipated to return.A follow-up report will be provided upon completion of investigation.In accordance to 21 cfr 803.56, if we obtain additional information, which was not known or was not available when the initial report was submitted, then the supplemental report will be submitted to the fda.There is no report of serious injury or death associated with this event.
 
Event or Problem Description
"the grasper was being used in a adrenalectomy.Staff noticed blue pieces from mesh in the patient which they retrieved." patient status: patient is ok.
 
Additional Manufacturer Narrative
The event unit was returned for evaluation.Upon inspection, engineering observed that the pad and jaw of the event unit were damaged.Testing confirmed that the damage observed to the event unit was caused by ultrasonic instrumentation.The root cause of the damage is use error, as the device is not designed to withstand direct contact from ultrasonic instrumentation.Applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C4119, 32CM GRASPER REPOS CART, 10/BX
Common Device Name
NWV
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key6115089
Report Number2027111-2016-00777
Device Sequence Number1360543
Product Code NWV
UDI-Device Identifier00607915116859
UDI-Public(01)00607915116859(17)190706(30)01(10)1274388
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/06/2019
Device Model NumberC4119
Device Catalogue Number100863901
Device Lot Number1274388
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/26/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/18/2016
Supplement Report FDA Received Date05/09/2017
Patient Sequence Number1
-
-