BAXTER HEALTHCARE - SWINFORD OUTLET PORT CLAMP; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number SPC4527
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/25/2016
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during continuous ambulatory peritoneal dialysis (capd) therapy, the patient had a leak when they were using the outlet port clamp.The patient stated that they clamped the drain line, but noted that even though they have clamped the line, solution was still leaking out.The clamp was new and the patient went to use the clamp at the end of dialysis.After clamping the line, the fluid was seeping through where the clamp was and had dripped out the end onto work surface.As a result, the patient used multiple clamps to pinch the line.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: the actual device and companion samples were received for evaluation.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional testing was performed with no issues noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OUTLET PORT CLAMP
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - SWINFORD
swinford
Manufacturer (Section G)
BAXTER HEALTHCARE - SWINFORD
foxford road
ni
swinford
EI  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6117122
MDR Text Key60548602
Report Number1416980-2016-17499
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/14/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/30/2020
Device Catalogue NumberSPC4527
Device Lot Number15J01H80
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received11/24/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
TWIN BAGS
-
-