• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125325-23
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/25/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was returned for analysis.Visual, dimensional, and functional inspections were performed on the returned device.The reported difficulty removing the guide wire and collapsed lumen were not confirmed with inspection of the returned device.The investigation was unable to determine a conclusive cause for the reported difficulties.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that there was no resistance noted when advancing the 3.25x23 xience alpine stent delivery system (sds) over the non-abbott guide wire (gw).The xience alpine was pressurized one time at 18 atms and the stent was deployed without issues.However,as the xience alpine sds was pulled out of the deployed stent, the sion gw also came out unexpectedly, as the sds and gw were stuck together.Thus, the devices were removed out of the anatomy as a single unit.There was no resistance met between the deployed stent and the sds when the sds was removed.The physician comments that the gw lumen on the sds may have collapsed after the single inflation.No significant delay in the procedure was reported.No adverse patient effects were reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6117932
Report Number2024168-2016-08188
Device Sequence Number10945944
Product Code NIQ
UDI-Device Identifier08717648199455
UDI-Public(01)08717648199455(17)190214(10)6020341
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/14/2019
Device Catalogue Number1125325-23
Device Lot Number6020341
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2016
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/27/2016
Initial Report FDA Received Date11/21/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE WIRE: SION
-
-