• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-FRANCE XIA 3 TITANIUM ROD DIAM 6MM TI ALLOY L 600MM; PEDICLE SCREW SPINAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-FRANCE XIA 3 TITANIUM ROD DIAM 6MM TI ALLOY L 600MM; PEDICLE SCREW SPINAL SYSTEM Back to Search Results
Catalog Number 48233600
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Codes Injury (2348); No Code Available (3191)
Date of Event 10/27/2016
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that: on (b)(6) 2015, the patient underwent the surgery.After surgery, when the surgeon checked x-ray, the rod was broken(l5-s).Therefore, the patient underwent the revision surgery on (b)(6) 2016.During revision surgery, the surgeon tightened the blocker using the torque wrench.When the surgeon pulled out the torque wrench, the screw head separated(l5 right and il left).Therefore the surgeon changed the screw to another same size screw.
 
Additional Manufacturer Narrative
Method: device history review; complaint history review; risk assessment; results: manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.Conclusion: no further investigation for this event is possible at this time as no devices and insufficient information was received.
 
Event or Problem Description
It was reported that: on (b)(6) 2015, the patient underwent the surgery.After surgery, when the surgeon checked x-ray, the rod was broken(l5-s).Therefore, the patient underwent the revision surgery on (b)(6) 2016.During revision surgery, the surgeon tightened the blocker using the torque wrench.When the surgeon pulled out the torque wrench, the screw head separated (l5 right and il left).Therefore the surgeon changed the screw to another same size screw.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA 3 TITANIUM ROD DIAM 6MM TI ALLOY L 600MM
Common Device Name
PEDICLE SCREW SPINAL SYSTEM
Manufacturer (Section D)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR  33610
MDR Report Key6118595
Report Number0009617544-2016-00469
Device Sequence Number10878798
Product Code MNH
UDI-Device Identifier04546540560735
UDI-Public(01)04546540560735
Combination Product (Y/N)N
PMA/510(K) Number
K071373
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2016
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number48233600
Device Lot NumberJLY
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 10/27/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/21/2016
Supplement Report FDA Received Date03/10/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-