STRYKER NEUROVASCULAR CORK TARGET 360 SOFT 3MM X 6CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0035473060
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/26/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device is not available.
 
Event or Problem Description
It was reported that during coil embolization procedure the coil prematurely separated at the detachment zone.The physician safely removed the coil from the patient's body with a retrieval device without clinical consequence to the patient.
 
Additional Manufacturer Narrative
Expiration date/manufacturing date: added.The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.The subject device was not returned; therefore, physical as well as a functional evaluation could not be performed.Information available indicated that the device was confirmed to be in good condition prior to use and was prepared ad per the device direction for use (dfu).Based on the information currently available the exact cause for the reported issue cannot be determined.
 
Event or Problem Description
It was reported that during coil embolization procedure the coil prematurely separated at the detachment zone.The physician safely removed the coil from the patient's body with a retrieval device without clinical consequence to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 SOFT 3MM X 6CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key6119355
Report Number3008881809-2016-00273
Device Sequence Number52995
Product Code HCG
UDI-Device Identifier04546540676429
UDI-Public(01)04546540676429(17)190531(10)19079427
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K102672
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Catalogue NumberM0035473060
Device Lot Number19079427
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 10/26/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/21/2016
Supplement Report FDA Received Date02/06/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR SL-10 MICROCATHETER (STRYKER)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient Weight98
-
-