| Model Number |
H7493808008320 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/19/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The balloon was tightly folded.There was contrast in the inflation lumen.There was blood in between balloon folds.The outer shaft, inner shaft, balloon and tip were microscopically examined.The hypotube shaft was completely separated 52.5 cm from the hub.The fracture faces were oval as if kinked prior to separation.There were numerous hypotube kinks.There was tip damage.Inspection of the remainder of the device presented no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 01-nov-2016.It was reported that crossing difficulties were encountered.The 90% stenosed target lesion was located in a moderately tortuous and severely calcified coronary artery.A 3.25mm x 8mm quantum¿ maverick¿ was advanced but failed to cross the lesion.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed the hypotube break.
|
| |
|
Search Alerts/Recalls
|