Catalog Number 415108
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Two used samples were received for evaluation.The samples were tested for iso luer taper, weepage, flow, and pressure according to the specifications with passing results.No leakage was observed.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or nonconformances of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number or lot number.Based on the results of this investigation, no specific conclusions can be made regarding the cause of the reported event.The returned set met requirements according to specification, and the reported failure could not be reproduced.If additional pertinent information becomes available a follow-up report will be filed.Note: this case is being filed retrospectively as a result of a review of recent customer complaint information.Based on additional information and details provided in another complaint case, it was determined that this case is reportable in accordance with the requirements of 21 cfr 803.B.Braun has conducted a retrospective review for all complaints of a similar nature in accordance with internal procedure (b)(4).
 
Event or Problem Description
As reported by user facility: event 1: leaking both blood and saline between the introcan and luer lock of the safeday.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDAY¿
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
MDR Report Key6120354
Report Number2523676-2016-00795
Device Sequence Number212179
Product Code FPA
UDI-Device Identifier04046964296215
UDI-Public(01)04046964296215(17)170430(10)0061372080
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K955585
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 11/19/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2017
Device Catalogue Number415108
Device Lot Number0061372080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2014
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2014
Initial Report FDA Received Date11/22/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-