| Catalog Number |
V7490 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/10/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as event three of b.Braun medical inc.Internal report # (b)(4).The actual device involved in the reported incident was not returned for evaluation, however two pictures were provided with the report.The pictures were evaluated, and it appears that there was blood leakage near the bonded joint of the non-vented spike assembly.Without the actual sample a thorough sample analysis could not be performed and no specific conclusions can be drawn.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or nonconformances of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number or lot number.If additional pertinent information becomes available a follow-up report will be filed.Note: this case is being filed retrospectively as a result of a review of recent customer complaint information.Based on additional information and details provided in another complaint case, it was determined that this case is reportable in accordance with the requirements of 21 cfr 803.B.Braun has conducted a retrospective review for all complaints of a similar nature in accordance with internal procedure cop-qp-8000051.
|
| |
|
Event or Problem Description
|
|
As reported by user facility: event 3.Iv tube is leaking blood during transfusion.
|
| |
|
Search Alerts/Recalls
|