ARTHREX, INC. IBAL TKA,BEARING IMP,PS,SZ 3,14MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
AR-513-BD14 |
| Health Effect - Clinical Codes |
Pain (1994); Loss of Range of Motion (2032); Swelling (2091)
|
| Date of Event |
10/31/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No further patient information was provided at the time of this report or made available in response to arthrex's follow-up requests.The evaluation revealed that the returned implant has normal wear patterns on the articulating surface.There are gouges and irregular surface textures on the a surface.There are scrape marks on the post and burnishing marks on the c surface.The damages observed on the locking tabs and post were most likely caused during the extraction of the implant.At the current time the cause of the event cannot be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2015 a patient underwent a right tka procedure.On (b)(6) 2016 the surgeon performed a revision right tka due to patient pain.During the revision procedure the surgeon explanted the tibial tray ar-503-tttd, lot 1307267 (cc99503 line 174688), femoral implant ar-516-4r, lot 108761424 (cc99503 line 174689), bearing implant ar-513-bd14, lot 113601340 (cc99503 line 174690) and patella implant ar-504-psb8, lot 113601323 (cc99503 line 174813).To complete the procedure the surgeon used another manufacturer's product.Patient was a female, age (b)(6).Patient medical records dated (b)(6) 2016 noted examination of the right knee found mild effusion of the right knee.Right knee demonstrated no pain associated with motion in any direction.Mild mid-flexion laxity was noted.Patient has pain in the knee that is increased with activity and weight bearing, as well as interference with activities of daily living.Pain through a limited passive range of motion with crepitus and swelling.X-ray records also noted periarticular osteophytes and joint space narrowing.
|
| |
|
Search Alerts/Recalls
|
|
|