• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PRODIGY IPG; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PRODIGY IPG; SCS IPG Back to Search Results
Model Number 3799
Device Problem Device Stops Intermittently (1599)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.(b)(4).
 
Event Description
It was reported the patient's (b)(4) ipg turned off randomly, which in turn resulted in loss of stimulation.Diagnostics indicated normal lead impedance values.Reprogramming was attempted to no avail.Surgical intervention is planned to address the issue.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Follow-up identified the patient underwent surgical intervention to explant and replace the ipg.Stimulation was restored post-operatively.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODIGY IPG
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key6120667
MDR Text Key60657834
Report Number1627487-2016-05813
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 01/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2016
Device Model Number3799
Device Lot Number4914777
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/03/2016
Initial Date FDA Received11/22/2016
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received12/27/2016
01/12/2017
01/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age55 YR
-
-