• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Imprecision (1307)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/25/2016
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative inspected the imaging system on-site.The reported issue could not be replicated during inspection.A full navigation system check-out was completed and all tests passed.Full system functionality was confirmed and the system was returned to service.
 
Event Description
A site representative reported that during a cranial resection procedure, the navigation system was inaccurate.It was reported that the patient reference with fiducials was not accurate.Navigation was discontinued because of this issue.There was a reported delay to the procedure of less than 1 hour due to this issue.The procedure was completed successfully without the use of navigation, and the patient was not impacted.No additional details were provided.
 
Manufacturer Narrative
The software investigation found that the reported event was unrelated to a software issue.Image analysis performed found that the fiducials were placed on the back of the head and localized.It was recommended to the site that the fiducials be spread out.The software functioned as designed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
826 coal creek circle
louisville, CO 80027
7208902439
MDR Report Key6121335
MDR Text Key60693486
Report Number1723170-2016-05497
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450968
UDI-Public00613994450968
Combination Product (y/n)N
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 03/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/22/2016
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733858
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/27/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/13/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age64 YR
Patient Weight140
-
-