Model Number 3CX*FX15RW30 |
Device Problem
Device Alarm System (1012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/03/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
Terumo has not received the actual device; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow up report when the investigation is complete and more information becomes available.(b)(4).Conclusion not yet available-evaluation in progress.
|
|
Event Description
|
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass the bubble detector alarmed on their sorin s5 heart-lung machine.They did not find any air so they reset and went back on pump.Through the rest of the case they saw air activity being displayed on the heart-lung machine.Case was 1.5 to 2 hours long.Sensitivity was set to "medium".*no known impact or consequence to patient.*product was not changed out.*procedure completed successfully.
|
|
Manufacturer Narrative
|
A sample was not returned for evaluation and the lot number of the affected product was not provided; therefore, a complete investigation could not be performed and a definitive root cause could not be determined.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|