• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORAX MEDICAL, INC. FENIX CONTINENCE RESTORATION SYSTEM; IMPLANTED FECAL INCONTINENCE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. FENIX CONTINENCE RESTORATION SYSTEM; IMPLANTED FECAL INCONTINENCE DEVICE Back to Search Results
Model Number FXS19
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Wound Dehiscence (1154); Incontinence (1928); Perforation (2001)
Date of Event 10/24/2016
Type of Reportable Event Serious Injury
Event or Problem Description
Following a surgical procedure for reinforcement of the anal sphincter due to fecal incontinence, a patient experienced fecal incontinence (lack of effect) leading to fenix device explant.The fenix device was used as part of the surgical procedure.Uneventful surgical procedure and device implant on (b)(6) 2016.Uneventful device explant on (b)(6) 2016.Device was found in correct position/geometry at the time of explant.
 
Additional Manufacturer Narrative
-addition to include disability or permanent damage now that reason for removal is confirmed as unhealed perforation during implant procedure.-addition to include vaginal wall perforation and suture repair during implant surgery on (b)(6) 2016.-addtion to include device protruding through suture repair site.(b)(4).
 
Event or Problem Description
Following a surgical procedure for reinforcement of the anal sphincter due to fecal incontinence, a patient experienced a perforation of the vaginal wall during implant which was still present at follow up visit leading to fenix device explant.The fenix device was used as part of the surgical procedure.-surgical procedure and device implant on (b)(6) 2016 that included a posterior perforation of the vaginal wall during dissection which was sutured shut before the device implant.-a follow up visit noted the device protruded through the posterior vaginal wall at the site of the suture repair and an uneventful device explant occurred on (b)(6) 2016.-device was found in correct position/geometry at the time of explant.
 
Additional Manufacturer Narrative
Include disability or permanent damage now that reason for removal is confirmed as unhealed perforation during implant procedure.Include vaginal wall perforation and suture repair during implant surgery on (b)(6) 2016.Include device protruding through suture repair site.Follow up report 2 included the removal of "device found in correct position/geometry at the time of explant".
 
Event or Problem Description
Following a surgical procedure for reinforcement of the anal sphincter due to fecal incontinence, a patient experienced a perforation of the vaginal wall during implant which was still present at follow up visit leading to fenix device explant.The fenix device was used as part of the surgical procedure.-surgical procedure and device implant on (b)(6) 2016 that included a posterior perforation of the vaginal wall during dissection which was sutured shut before the device implant.-a follow up visit noted the device protruded through the posterior vaginal wall at the site of the suture repair and an uneventful device explant occurred on (b)(6) 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FENIX CONTINENCE RESTORATION SYSTEM
Common Device Name
IMPLANTED FECAL INCONTINENCE DEVICE
Manufacturer (Section D)
TORAX MEDICAL, INC.
4188 lexington avenue n
shoreview MN 55126
MDR Report Key6123053
Report Number3008766073-2016-00097
Device Sequence Number1154965
Product Code PMH
UDI-Device Identifier00855106005523
UDI-Public00855106005523
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
H130006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Physician
Device Expiration Date02/04/2020
Device Model NumberFXS19
Device Lot Number10061
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 10/28/2016
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date11/22/2016
Supplement Report FDA Received Date01/03/2017
04/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Disability;
Patient Age69 YR
-
-