• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS DE MEXICO POWERFLEXPRO 8MM2CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS DE MEXICO POWERFLEXPRO 8MM2CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 4400802S
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/28/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A device history record review was performed and showed that these lots of products met all requirements per the applicable manufacturing quality plan.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
It was reported that a powerflex pro balloon was inflated in the lesion but it ruptured at its nominal pressure.It was unknown how the procedure completed but the procedure was completed successfully.There was no reported patient injury.The target lesion was unknown.The target lesion was unknown.The patient¿s vessel level of tortuousness and calcification was unknown.The rate of stenosis was unknown.The product was clinically used and will not be returned for analysis.
 
Additional Manufacturer Narrative
An 8x20mm powerflex pro balloon was inflated in the lesion but it ruptured at its nominal pressure.It was unknown how the procedure completed but the procedure was completed successfully.There was no reported patient injury.The target lesion was unknown.The target lesion was unknown.The patient¿s vessel level of tortuousness and calcification was unknown.The rate of stenosis was unknown.The device was not returned for analysis.A device history record (dhr) review of lot 17377043 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics were not provided.According to the instructions for use, ¿the rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.¿ neither the dhr nor the information available suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERFLEXPRO 8MM2CM 80
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS DE MEXICO
circuito int nte 1820
juarez 32575
MX  32575
MDR Report Key6123437
Report Number9616099-2016-00746
Device Sequence Number13741899
Product Code LIT
UDI-Device Identifier20705032059777
UDI-Public20705032059777
Combination Product (Y/N)N
PMA/510(K) Number
K112797
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 01/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date10/31/2018
Device Model Number4400802S
Device Catalogue Number4400802S
Device Lot Number17377043
Was Device Available for Evaluation? No
Distributor Facility Aware Date10/28/2016
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 10/28/2016
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/22/2016
Supplement Report FDA Received Date01/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-