| Catalog Number |
6430530 |
| Medical Device Problem Codes |
Failure to Align (2522); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/27/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.
|
| |
|
Event or Problem Description
|
|
Levels implanted- l4-5 pre-operative diagnosis: spinal canal stenosis it was reported that the patient underwent a fusion surgery, the set screw rotated idly and it could not be broken off at final tightening.Product came in contact with the patient.Patient complications were reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: macroscopic and optical examination revealed thread crest/flank damage; this damage appears to have initiated near the start of the thread, and is evident around the damaged portion of the thread.Functional evaluation with a sample mas head found the set screw was unable to be fully engaged.The above observations are consistent with misalignment of the mas and set screw threads during construct assembly.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|