Catalog Number 6430530
Medical Device Problem Codes Failure to Align (2522); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/27/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.
 
Event or Problem Description
Levels implanted- l4-5 pre-operative diagnosis: spinal canal stenosis it was reported that the patient underwent a fusion surgery, the set screw rotated idly and it could not be broken off at final tightening.Product came in contact with the patient.Patient complications were reported as unknown.
 
Additional Manufacturer Narrative
Product analysis: macroscopic and optical examination revealed thread crest/flank damage; this damage appears to have initiated near the start of the thread, and is evident around the damaged portion of the thread.Functional evaluation with a sample mas head found the set screw was unable to be fully engaged.The above observations are consistent with misalignment of the mas and set screw threads during construct assembly.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key6123869
Report Number1030489-2016-03229
Device Sequence Number783717
Product Code NKB
UDI-Device Identifier00643169545281
UDI-Public00643169545281
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K143375
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2023
Device Catalogue Number6430530
Device Lot Number0409995W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2016
Initial Date Received by Manufacturer 10/27/2016
Supplement Date Received by ManufacturerNot provided
01/19/2017
Initial Report FDA Received Date11/23/2016
Supplement Report FDA Received Date02/07/2017
09/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-