• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN DOMINICAN REPUBLIC INC. SAFEDAY¿; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN DOMINICAN REPUBLIC INC. SAFEDAY¿; SET, ADMINISTRATION, INTRA Back to Search Results
Catalog Number 352899
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as event two of b.Braun inc.(b)(4).Twelve unused samples were received for evaluation.All twelve sets were tested for leakage according to the specification with passing results.No leakage was observed from any bonded joint or safeday y-site on any set.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or nonconformances of this nature.No adverse quality trends of this nature were identified during the complaint review process for the reported catalog number or lot number.Based on the results of this investigation, no specific conclusions can be made regarding the cause of the reported event.The returned set met requirements according to specification, and the reported failure could not be reproduced.If additional pertinent information becomes available a follow-up report will be filed.Note: this case is being filed retrospectively as a result of a review of recent customer complaint information.Based on additional information and details provided in another complaint case, it was determined that this case is reportable in accordance with the requirements of 21 cfr 803.B.Braun has conducted a retrospective review for all complaints of a similar nature in accordance with internal procedure (b)(4).
 
Event or Problem Description
As reported by user facility: event 2: blood backing up into the tubing.The blood was leaking from the middle y-site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDAY¿
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km22 autopista las americas
santo domingo, dominican republic
DR 
MDR Report Key6124474
Report Number9614279-2016-00143
Device Sequence Number19035751
Product Code FPA
UDI-Device Identifier04046964181931
UDI-Public(01)04046964181931(17)170831(10)0061395513
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K955585
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 12/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2017
Device Catalogue Number352899
Device Lot Number0061395513
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/23/2014
Initial Report FDA Received Date11/23/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-